Development of a robust TaqMan probe-based one-step multiplex RT-qPCR for simultaneous detection of SARS-CoV-2 and Influenza A/B viruses

Abstract

Background

During the coronavirus disease 2019 (COVID-19) pandemic, the simultaneous detection of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) and Influenza A, and Influenza B viruses is essential for rapid differential diagnosis in patients with similar symptoms, especially during “flu season” in the post-pandemic era. So far, several multiplex methods have been approved for the simultaneous detection of SARS-CoV-2, Influenza A, and Influenza B. However, due to the rapid mutation rate of the SARS-CoV-2 genome and the emergence of new variants, existing methods must be improved and updated.

Methods

To identify a highly conserved region in the SARS-CoV-2 N-gene, a genomic survey was performed to increase the sensitivity and specificity of primer and probe sets targeting the SARS-CoV-2 genome. The 95% LLOD (95% lower limits of detection) were calculated by probit analysis. A total of 70 predetermined clinical samples using singleplex RT-qPCR assays, were included. The clinical performance of the multiplex RT-qPCR assay was determined and compared with a commercial multiplex kit. The Cohen’s kappa coefficient, P-value (McNemar’s test), Passing-Bablok regression, and Bland Altman agreement analysis were determined to monitor the agreement of the assays.

Results

The novel SARS-CoV-2 primer and probe set designed in this assay was able to detect all variants of concern (VOCs) and variants of interest (VOIs) with high analytical and clinical performance. The 95% LLOD for the multiplex RT-qPCR was 20 copies per reaction for the N gene of SARS-CoV-2, 2 copies per reaction for M1 gene of Influenza A and NS1 gene of Influenza B. The diagnostic sensitivity of the multiplex RT-qPCR was 94.4%, 93.7%, and 100% for the detection of SARS-CoV-2, Influenza A, and Influenza B genomes, respectively. Moreover, the specificity was identical (100%) in both assays. According to the agreement analysis results, there was no statistical difference between our multiplex assay and the commercial kit.

Conclusions

In this study, we developed a novel in-house made multiplex RT-qPCR assay, with high sensitivity, specificity, and reliability for the diagnosis of SARS-CoV-2 infection in clinical samples. This is valuable during Influenza seasons when influenza co-circulates with SARS-CoV-2, as it saves costs, time, and thus specific and timely treatment of patients.

Peer Review reports

Introduction

SARS-CoV-2, Influenza A, and Influenza B are significant global health threats with the potential of causing economic, medical, and severe public health crises with millions of infections and deaths throughout the world. These viruses are the most important human pathogens and usually cause mild upper respiratory diseases that spread rapidly among people [1,2,3,4,5]. We now critically needed an accurate and readily available molecular diagnostic technique to screen and identify patients infected with these pathogens. As a significant trait that influences their infectivity and COVID-19 pandemic patterns, Influenza viruses have been known to be important causative agents of co-infection in patients, which rise mainly in the flu season. Viral co-infections with COVID-19 disease are known as a major cause of morbidity and mortality and can pose a challenge to healthcare providers because they have very similar clinical features [6,7,8,9]. However, individuals infected with SARS-CoV-2 have a longer incubation period (2–14 days) until the onset of clinical symptoms compared to individuals infected with influenza A and B [10]. In light of this fact, the design and development of an in-house made one-step multiplex RT-qPCR can help accurately diagnose and differentiate SARS-CoV-2 from other respiratory viruses to manage COVID-19 patients and reduce time and cost. Multiplex RT-qPCR is considered one of the best methods for the detection of SARS-CoV-2 and Influenza viruses in clinical samples using multiple primer and probe sets that bind different targets to specifically amplify it [11,12,13]. In this study, we aim to develop a robust method for a cheap and time-saving procedure to detect and differentiate SARS-CoV-2 and Influenza A and B, and consequently, specific and timely treatment of patients, and also highlighted the pitfalls of the one-step multiplex RT-qPCR assay. For this purpose, SARS-CoV-2 genomic survey was performed to increase the sensitivity and specificity of the primer and probe binding to the SARS-CoV-2 genome.

Methods

In-silico studies for primer and probe set design

Nucleotide sequences of the SARS-CoV-2 N-gene (~ 43,000 sequences) were obtained from the NCBI database (https://0-www-ncbi-nlm-nih-gov.brum.beds.ac.uk/sars-cov-2/, accessed on March 2022) and aligned to identify conserved and non-conserved regions. Usegalaxy server (https://usegalaxy.org), Aliview software version 1.26, and MEGA10 software were used for sequence editing and minor manual adjustments. The selected sequences included Omicron BA.3 (78 sequences), Theta (42 sequences), and the Wuhan reference sequence (40,000 sequences). There were 200 sequences for variants including Alpha, Beta, Gamma, Delta, Omicron BA.1, Omicron BA.2, Omicron BA.4, Omicron BA.5, Epsilon, Zeta, Eta, Iota, Kappa, and Lambda (Table 1). To gain insight into the alignment heterogeneity, we calculated the positional nucleotide numerical summary (PNNS) and entropy values (H(i)) by using the Alignment Explorer server (http://www1.szu.cz:8080/EntropyCalcWeb). The entropy plot was made by using Microsoft Excel. Specific primer and probe set for the detection of SARS-CoV-2, Influenza A, and Influenza B are shown in Table 2. When designing and selecting multiple PCR primer and probe sets for multiplex assay, it is important to consider the compatibility between primer and probe sets, such as dimerization of oligonucleotides or secondary structures, GC content, primer and probe sets Tm, amplicon length, and oligonucleotides specificity. The IDT OligoAnalyzer™ tool (https://www.idtdna.com/pages/tools/oligoanalyzer), Oligo (version 7.60), and AlleleID 6 software were used for oligonucleotides analysis. The presence of secondary structures in primer and probe binding sites can cause poor amplification efficiency; therefore we designed the primer and probe to hybridize to the cyclic structure. Secondary structure and primer-probe binding sites of the N gene predicted by the RNAfold web server (http://rna.tbi.univie.ac.at/cgi-bin/RNAWebSuite/RNAfold.cgi). We also aligned the N-gene of SARS-CoV-2 with other coronaviruses such as SARS-CoV-1, MERS-CoV, HCoV-229E, HCoV-NL63, HCoV-HKU1, and HCoV-OC43 to identify conserved and non-conserved regions among coronaviruses. A simple workflow for the design and development of our multiplex assay is shown in Fig. 1.

Table 1 Selected sequences related to variants of SARS-CoV-2 for multiple alignments
Table 2 Primers and probes used in our multiplex RT-qPCR
Fig. 1
figure 1

A simple workflow for designing and developing of the multiplex assay. For these purposes, (1) in silico approaches, such as gene target selection, multiple alignments, sequences processing and primer-probe sets design, and (2) experimental methods, such as real time PCR optimization, and analytical and clinical evaluation of the assays are crucial

Construction of positive control plasmids

The SARS-CoV-2 N, Influenza A M1, and Influenza B NS1 amplicons were amplified by RT-PCR from positive clinical samples. Amplicons were cloned into the pUCM-T vector using the BioBasic TA Cloning kit (BioBasic, Canada) according to the manufacturer’s instructions to obtain the appropriate plasmids. Purified plasmids were sequenced to confirm sequence accuracy (Supplemental Information 1) and were also used to test for their sensitivity and specificity in the conventional PCR and to derive standard curves for the multiplex RT-qPCR assay. Copy number (copies/µL) was calculated using the following equation: [C (ng/µL)*6.022 × 1023/ N*660 (g/mol)*109 (ng/g)], in which C represents the concentration of plasmid (ng/µL); N is the length of the plasmid (number of nucleotides), and 660 is the average mass of 1 bp dsDNA [14].

SYBR Green real-time PCR

When designing multiplex RT-qPCR assays, it is critical to choose primers that are highly specific and do not produce primer dimers. Before multiplexing, all primer sets were tested under singleplex, duplex, triplex and quadruplex conditions using the SYBR Green RT-qPCR assay. The SYBR Green RT-qPCR assay was performed using the ABI StepOnePlus™ Real-Time PCR System (Applied-Biosystems) in the presence of 2X SYBR-Green PCR Master Mix (Sinacolon, Iran). We carefully optimized the real-time PCR conditions to minimize primer dimer formation by optimizing the primer concentration and the temperature using the melting curve analysis. Amplifications were performed in a final volume of 20µL, including 5µL of positive control plasmids as a template. Cycling conditions were optimized according to the amplicon size and the Tm of primers, ending with a melting curve from 60 to 90 °C. Fluorescence was measured at the end of each cycle. After optimization, we performed SYBR Green real-time PCR amplification using the quantified SARS-CoV-2 positive control plasmid. Standard curves were generated by plotting Cycle threshold (Ct) versus plasmid copy number. Linear regression analysis was carried out for the SARS-CoV-2 N gene.

TaqMan probe-based multiplex RT-qPCR assay condition

Multiplex RT-qPCR was optimized for all reactions and performed using the one-step RT-qPCR Master Mix (Pishtazteb co, Tehran, Iran), in ABI StepOnePlus™ Real-Time PCR System (Applied-Biosystems). The reaction mixture (final volume: 20 µL) consists of 9 µL of RT-qPCR solution, 0.225 µM of each SARS-CoV-2 and Rnase P forward and reverse primers, 0.8 µM of the SARS-CoV-2 probe, 0.15 µM of each Influenza A and Influenza B primers and probes sets, 0.5 µM of the probe for RNase P and 10µL of RNA template. The RNase P gene was used as an internal control for RNA extraction, sampling, and RT-qPCR process monitoring to avoid false-negative results. To monitor cross-dimer or contamination, we added nuclease-free water to negative control tubes. Thermal cycling conditions were performed as follows: 20 min at 50 °C for reverse transcription, 3 min at 95 °C for PCR initiation activation, and 40 cycles of 95 °C for 10s (denaturation) and 55 °C for 35s (annealing/extension).

LLOD (lower limits of detection) of the multiplex RT-qPCR assay

After optimizing primer and probe sets under different conditions (singleplex to multiplex) using positive control plasmids, we generated standard curves for all target genes. After diluting the plasmids in a 10-fold ratio (ranging from 3 × 105, 3 × 104, 3 × 103, 3 × 102, 3 × 101, 3 × 100, and 0.3 copies per reaction), standard curves were generated by plotting Cycle threshold (Ct) versus plasmid copy number and linear regression analysis were carried out for the N, M1, and NS1 gene targets. The analytical sensitivity of the multiplex RT-qPCR assay was measured by testing 10-fold serial dilutions of quantified standard plasmids containing SARS-CoV-2, Influenza A, and Influenza B gene fragments. Each dilution was tested in triplicate by the multiplex RT-qPCR assay. The 95% lower limits of detection (95% LLOD) were calculated using probit analysis. For the determination of 95% LLOD, each dilution was tested in 6 replicates on two independent runs, and the lower limit of detection (LLOD) was defined as the concentration of copies/reaction of the lowest dilution that could be detected with 95% probability.

The precision of the multiplex RT-qPCR assay

Repeatability (intra-assay precision) and reproducibility (inter-assay precision) of the multiplex RT-qPCR assay were determined using three different concentrations (5 × 105, 5 × 103 and 5 × 101 copies per reaction) of each plasmid standard. For intra-assay, each dilution was analyzed in triplicate in one reaction, while for inter-assay; each dilution was analyzed in three independent reactions on day 1, day 2, and day 3. Intra- and inter-assay were calculated for each dilution and expressed as % coefficients of variation (%CV).

Analytical specificity

We evaluated the analytical specificity of the multiplex assay by using common respiratory pathogens, including Influenza A virus, Influenza B virus, SARS-CoV-2, Respiratory Syncytial Virus (A and B), Rhinovirus, Adenovirus (B and C) and Epstein-Barr virus (EBV).

Evaluation of clinical performance

For evaluation of the clinical performance, a total of 70 archived predetermined respiratory swab specimens were subjected to the multiplex RT-qPCR assay. The samples were previously tested for SARS-CoV-2, Influenza A, and Influenza B by singleplex RT-qPCR diagnostic assays as the gold standard. The singleplex RT-qPCR tests were performed using the one-step RT-qPCR kit (Pishtaz Teb Diagnostics, Tehran, Iran), according to the manufacturer’s instructions. To validate the outcome of the multiplex RT-qPCR assay, we compared our results and tested the clinical samples with Viga SARS-CoV-2 and Influenza A/B molecular diagnostic kit (ROJE Technologies, Iran). Out of these samples, 18 were positive for SARS-CoV-2, 16 were positive for influenza A, 10 were positive for influenza B, and 26 were negative for SARS-CoV-2 and influenza A and B. The specimens were placed into sterile tubes containing 3ml of viral transport media (VTM) consisting of Hank’s balanced salt solution at pH 7.4 containing BSA (1%), amphotericin (15 µg/mL), penicillin G (100 units/mL), and streptomycin (50 µg/mL). Viral RNA was extracted from 200 µL of the VTM by using an RNA extraction kit (RNJia kit, ROJE Technologies, Iran), and eluted in 60 µL of nuclease-free water. RNA extracts were stored at -70 °C for further analyses. The agreement between our multiplex assay and the commercial multiplex kit was evaluated using Cohen’s kappa index, McNemar’s test (P-value), Passing-Bablok regression, and Bland Altman agreement analysis. Passing-Bablok regression was used to compare correlations Ct values in clinical samples between our multiplex assay and commercial multiplex kit. Bland-Altman analysis was used to determine the bias and limits of agreement between the two assays, corresponding to the 95% CI of the mean bias of all paired measurements.

Testing of co-infections

To evaluate the performance of the current assay in cases with co-infections, we included and combined different viruses (containing of: SARS-CoV-2, Influenza A and Influenza B) with different Ct values. We prepared a total of 8 co-infected samples for SARS-CoV-2 and Influenza A and B viruses.

Statistical analysis

Statistical analysis including mean, standard deviation (SD), and coefficient of variation (CV %) for Ct values, box plot, and creating of entropy plot were performed using Microsoft Excel. Statistical analysis, including Probit regression analysis (at the 95% probability level to determine the detection limits), Passing-Bablok regression, and Bland Altman agreement were performed with MedCalc Statistical Software version 22.007 (MedCalc Software Ltd, Ostend, Belgium; https://www.medcalc.org). McNemar’s test and the kappa index were used to analyze the statistical difference and agreement between the methods by SPSS Version 22 (IBM Corp., USA). A P-value of < 0.05 was considered statistically significant.

Results

Primer and probe sets were specific to SARS-CoV-2, Influenza A, and Influenza B

In the first of the COVID-19 pandemic, all nucleotide sequences of the SARS-CoV-2 were obtained from GenBank that belong Wuhan strain. During the waves of the pandemic and the continuously evolving nature of SARS-CoV-2, we added available nucleotide sequences of the variants (Alpha, Beta, Gama, Delta, Omicron BA.1, Omicron BA.2, Omicron BA.2.12.1, Omicron BA.4, Omicron BA.5). The variability analysis of N gene sequences using positional nucleotide numerical summary calculation (PNNS) and entropy plot (Fig. 2a) shows that a highly conserved region (802–905 bp) was selected as SARS-CoV-2 primer and probe annealing site. The compatibility analysis of primer and probe sets by in-silico approaches (IDT OligoAnalyzer™ and BLAST tools) showed some degree of homo- or hetero-dimerization. Therefore, we created primer/probe-template mismatches to reduce the possibility of cross-dimer formation between primer and probe sets. The mismatches are located at the 5′ ends of the N gene primers and probe.

Fig. 2
figure 2

The variability analysis of aligned sequences and prediction of secondary structure of the N primers and probe annealing sites. After “Positional nucleotide numerical summary calculation” the entropy values were calculated for the entire of the N gene. The entropy plot was obtained by plotting the entropy values against the N gene positions. The variation per position is expressed by the column height. The conserved region as amplicon is shown with red line (a). The primers and probe designed to hybridized with cyclic structure (no stem structure) (b)

The secondary structure of the N gene and the primer and probe binding sites are shown in Fig. 2b. Multiple sequence alignment confirmed that our proposed assay is specific for all subtypes of SARS-CoV-2, Influenza A, and Influenza B; no evidence of non–SARS-CoV-2, non- Influenza A, and non-Influenza B target matches was found (Fig. 3).

Fig. 3
figure 3

Alignment of SARS-CoV-2, Influenza A and Influenza B primer and probe sets with consensus sequences of SARS-CoV-2, Influenza A, and Influenza B subtypes

The multiple sequence alignment of SARS-CoV-2 N-gene with other human coronaviruses showed that primer and probe binding sites were conserved among the SARS-CoV-2 variants, but not among other human coronaviruses (Supplemental information 2).

Optimization primer sets using SYBR Green Real-time PCR

Before optimization and creation of mismatches in SARS-CoV-2-N primers, electrophoresis and melting curve analysis of the amplified N gene using conventional RT-PCR and SYBR Green real-time PCR showed the presence of a primer dimer (Supplemental information 3 (a)). After creating intentional mismatches and optimizing primer concentrations and cycling conditions under singleplex, duplex, triplex, and quadruplex conditions using SYBR Green real time-PCR, the primer dimer formation was mitigated. The melting curves obtained for all conditions are shown in supplemental information 3 (b-f). The curves showed only one peak corresponding to the amplification product, resulting in no primer dimmers. We also performed SYBR Green real-time PCR with N gene primers in duplicate using copy numbers of 3 × 106, 3 × 105, 3 × 104, 3 × 103, 3 × 102 and 30 per reaction. The amplification plots, standard curve, and R2 score are summarized in supplemental information 3 (g). A linear regression relationship was observed with an R2 score of 0.99 for the SARS-CoV-2 N gene.

The multiplex assay has a high analytical sensitivity

The results showed that the optimized SYBR Green real time-PCR using SARS-CoV-2-N primers under singleplex conditions can detect 30 copies per reaction of positive control plasmid. The analytical sensitivity of the singleplex assays for Influenza A and Influenza B was determined to be 5 copies per reaction [31 and 32].

The obtained standard curves (E, slope, and R2 score) and amplification plots for the N, M1, and NS1 genes using the multiplex assay are summarized in Fig. 4. The R2 scores were determined as 0.988 for the N, 0.992 for M1, and 0.920 for the NS1 gene. The analytical sensitivity of the multiplex assay on standard plasmids were 30 copies for the SARS-CoV-2, and 3 copies for both Influenza A and B, which corresponds to averaged Ct values of 34.84, 36.95, and 36.21 for N, M1, and NS1 genes, respectively (Supplemental information 4). These results were also in agreement with the singleplex real time-PCR assay. The 95% LLOD was determined for each gene in our multiplex RT-qPCR assay by probit analysis using positive control plasmid dilutions. The 95% LLOD was calculated to be 20 copies per reaction for the N gene and 2 copies per reaction for the M1 and NS1 genes.

Fig. 4
figure 4

The standard curves and amplification plots of our multiplex assay. A 10- fold serial dilution of plasmid standards containing cloned target sequences was prepared. The standard curves and amplification plots of the SARS-CoV-2-N gene (A), standard curves and amplification plots of the Influenza A-M1 gene (B) and standard curves and amplification plots of the Influenza B-NS1 gene (C)

Intra- and inter-assay variability

The results of intra- and inter-assays (Table 3) for the multiplex detection of N, M1, and NS1 genes revealed that the coefficient of variation (CV%) was all < 5%, which suggested the multiplex RT-qPCR assay is an accurate and reliable diagnostic tool for detection of the viruses.

Table 3 Intra- and Inter-assay in multiplex detection of SARS-CoV-2 (N gene), Influenza A (M1 gene), and Influenza B (NS1 gene)

The primer and probe sets had high analytical specificity among other respiratory viruses

The specificity tests showed that no amplification signals were detected using our multiplex assay for all four respiratory pathogens including respiratory syncytial virus (RSV), rhinovirus, adenovirus, and Epstein-Barr virus (EBV). We also observed no cross-reactivity among the three targets (SARS-CoV-2 and influenza A and B) within the multiplex assays (Supplemental information 5). These results suggest that our multiplex assay has good specificity for all target genes (N, M1, and NS1). The respiratory pathogens were isolated from clinical specimens and collected from subjects tested for these pathogens. In this study, mean Ct values for respiratory syncytial virus (RSV), rhinovirus, adenovirus, and Epstein-Barr virus (EBV) were, 21.53 ± 3.65, 22 ± 3.20, 24 ± 2.25, and 28 ± 2.25 respectively.

Our assay showed satisfactory clinical performance and agreement

The clinical performance of our proposed assay confirmed the high level of diagnostic sensitivity, specificity, accuracy, and agreement with reference singleplex RT-qPCR assays. Of the 18 SARS-CoV-2 samples, one tested negative, and of the 16 Influenza A samples, one tested negative by our multiplex assay. In contrast, of the 18 SARS-CoV-2 samples, two tested negative, and of the 16 Influenza A samples, one tested negative by the commercial multiplex kit. All 10 previously-tested Influenza B positive samples and 26 previously-tested negative samples were confirmed by both assays (Table 4) It should be noted that two clinical samples that were considered positive by the reference singleplex RT-qPCR assays were detected as negative by both multiplex methods. The samples with a Ct value of ≥ 37 are considered negative. The amplification curves of the clinical samples obtained by our multiplex assay are shown in supplemental information 6. A comparison of the Ct values ​​of each assay for the N, M1, and NS1 genes is shown in Fig. 5. A comparison of the various parameters of the multiplex assay and the commercial kit is provided in the supplemental information 7.

Table 4 Comparison of our multiplex assay results versus with commercial multiplex kit in clinical samples
Fig. 5
figure 5

Comparative analysis of Ct values between our multiplex assay and commercial multiplex kit using clinical samples. The average Ct value of the samples detected by our multiplex assay is slightly higher than the commercial multiplex kit. However, the results indicate that our primer and probe sets are favorable for simultaneous detection of SARS-CoV-2, Influenza A and Influenza B viruses

Using the clinical specimens previously tested by singleplex RT-qPCR assays, the sensitivity, specificity, accuracy, positive predictive values, and negative predictive values of our multiplex assay and the commercial multiplex kit were calculated (Table 5). The multiplex assay detected SARS-CoV-2, Influenza A, and Influenza B with sensitivity of 94.4%, 93.7%, and 100%, respectively. The sensitivity level of the commercial kit in detecting SARS-CoV-2, Influenza A, and Influenza B was 88.8%, 93.7%, and 100%, respectively. The assay specificity for our multiplex assay and the commercial multiplex kit was 100%. The accuracy of the multiplex assay in detecting SARS-CoV-2, Influenza A, and Influenza B was 97.7%, 97.7%, and 100%, respectively, while the accuracy of the commercial kit was 95.4%, 97.7%, and 100%.

Table 5 Summary of clinical performance of our multiplex assay and commercial multiplex kit compared to reference simplex assays

The agreement analysis results showed that there was no statistical difference between our multiplex assay and the commercial multiplex kit (McNemar’s test, P-value ˃ 0.05). Also, the Cohen’s kappa coefficient calculated in our multiplex assay with singleplex RT-qPCR assay, and the commercial kit with singleplex RT-qPCR assay was more than 94% and 90% respectively. So our multiplex had more agreements in detecting SARS-CoV-2, Influenza A, and Influenza B (Table 5).

The Passing-Bablok regression analyses between the Ct values of positive samples for SARS-CoV-2, Influenza A, and Influenza B obtained by both the multiplex assay and the commercial multiplex kit are shown in Fig. 6. There was a high correlation between Ct values obtained by the two assays for SARS-CoV-2 (R2 = 0.872 for the N gene), Influenza A (R2 = 0.874 for the M1 gene), and Influenza B (R2 = 0.827 for the NS1 gene). The Bland-Altman plots of between the Ct values obtained by the two assays on positive samples along with mean bias and their limits of agreement are shown in Fig. 6.

Fig. 6
figure 6

Summary of Passing-Bablok regression and Bland-Altman plots of Ct values obtained by our multiplex assay and commercial multiplex kit. Passing-Bablok regression and Bland-Altman plots for SARS-CoV-2 N gene (A), Influenza A M1 gene (B), Influenza B NS1 gene (C), and all genes (D). For Passing-Bablok regression plots, the dashed line indicates the ideal line, whereas the solid line shows the regression line of the distribution. For the Bland-Altman analyses, the solid line indicates the mean relative difference, and the dotted lines show the superior and inferior limits of agreement. Ct, cycle threshold

All the co-infected samples were detected using our multiplex assay

The results of the co-infected testing indicated that our multiplex assay can detect all target genes simultaneously. Calculated Ct values and amplification plots for 8 co-infected samples by our multiplex assay and commercial multiplex RT-qPCR kit are shown in supplemental information 8 and 9.

Discussion

The high prevalence of SARS-CoV-2, Influenza A, and Influenza B imposes a high financial burden on the healthcare system with millions of infections and deaths worldwide [15]. We now critically needed a fast, accurate and easily accessible diagnostic technique. The RT-qPCR assay allows amplification of nucleic acids using a primer and probe that bind to specific regions of the target viral genome to increase the assay specificity [10]. The multiplex RT-qPCR differs from the RT-qPCR in that it requires more than one set of primer and probe, which reduces the overall cost and time of the RT-qPCR assay [16]. Another advantage of multiplex RT-qPCR is the diagnosis and accurate differentiation of SARS-CoV-2 from other respiratory viruses such as Influenza A and B viruses, which manage COVID-19 patients with co-infections and secondary infections [17, 18]. Thus, the design and development of an in-house made one-step multiplex RT-qPCR method can help slow or stop the spread of these viruses. Two points were considered in the primer and probe sets design based on the genome of SARS-CoV-2 variants. The first was to identify the conserved and the unique region of the N-gene with the minimum entropy. The second was to recognize the non-stem structure of the N-gene, which can be easily accessed using primers and probe.

Researchers have designed RT-qPCR primer and probe for various targets in the SARS-CoV-2 genome, including the S, N, E, and RdRp genes [11, 19]. Our previous study shows that the N gene has higher specificity compared to the other SARS-CoV-2 genes and is also a better target for identifying new cases of SARS-CoV-2 in clinical samples, based on the real-time PCR Ct value. The multiplex RT-qPCR assay for simultaneous detection and differentiation of SARS-CoV-2 and other respiratory viruses has limitations related to the number of used primer and probe sets. Hence, we must use one set of primer and probe for SARS-CoV-2 detection [11]. In addition, the occurrence of mutations in the S, N, E, and RdRp genes is high, which can lead to primer and probe mismatches at annealing sites and increase false-negative result, especially for new variants with many mutations. Evaluation of the SARS-CoV-2 primer and probe sets of recommended by WHO and other studies showed that many mutations were located in annealing sites of most primers and probes [20,21,22,23].

Overall, a 104 bp region of the SARS-CoV-2 N-gene was identified using the RNAfold Web Server, which at 60 °C (RT-qPCR reaction temperature) had the lowest secondary structure and entropy compared to other regions of the genome (Fig. 2a-b). We also design for the first time a high coverage primer and probe set that were conserved among the SARS-CoV-2 variants as shown in Figs. 2 and 3, but were not conserved among other human coronaviruses.

The sensitivity of the multiplex RT-qPCR for detection of the SARS-CoV-2 and Influenza A and B genomes was consistent with previously reported ranges of 10–100 copies of RNA [24,25,26,27,28,29,30,31,32,33,34]. So, the sensitivity of our multiplex RT-qPCR assay is suitable for detection of SARS-CoV-2 and influenza A and B in clinical specimens.

It should be noted, that we determined repeatability and reproducibility of the multiplex RT-qPCR assay using three different concentrations of each plasmid standard. The coefficients of variation of the intra-assay repeatability and inter-assay reproducibility of the multiplex RT-qPCR assay were < 5.0% and < 4.0% respectively, (Table 3), demonstrating that our multiplex RT-qPCR assay is an accurate and reliable diagnostic tool for detection of the SARS-CoV-2, influenza A, and influenza B viruses.

The multiplex RT-qPCR assay specifically amplified SARS-CoV-2, influenza A, and influenza B RNA, but did not amplify nucleic acids from other respiratory pathogens such as respiratory syncytial virus (RSV), rhinovirus, Adenovirus, Epstein Barr virus, and other coronaviruses.

For evaluation of the clinical performance, in total 70 samples were provided, including 44 positive and 26 negative samples that were confirmed by the COVID-19 reference laboratory in Golestan province without knowing their RT-qPCR results (blind). Out of these samples, the 10 new samples isolated from pharyngeal swab specimens during the sixth wave of the pandemic (Omicron) were used for virus detection by both our multiplex RT-qPCR and commercial multiplex kit. It is revealed that two samples out of 70 did not match with the results of the multiplex RT-qPCR assay and three samples did not match with the results of the commercial multiplex kit. Therefore, the diagnostic sensitivity of our multiplex RT-qPCR was higher compared to the commercial multiplex kit. It should be noted that two samples out of 70 were similarly detected as false-negatives by both assays. These controversial results in the detection rate of the viruses are dependent on many factors, including the low viral load in clinical samples, degradation of RNA genome or repeated freezing and thawing of samples, RT-qPCR inhibitors, and specificity of primers and probes.

The Cohen’s kappa coefficient revealed that there was a more agreement between our multiplex RT-qPCR and the singleplex RT-qPCR assay (as gold standard) in the detection of SARS-CoV-2 (Table 5). In addition, the Passing-Bablok regression curves and Bland-Altman analysis between the Ct values of positive samples for SARS-CoV-2, Influenza A, and Influenza B revealed that there were high correlation between Ct values of our multiplex RT-qPCR and the commercial multiplex kit (Fig. 6). However, the average Ct value of the samples detected by our multiplex assay is slightly higher compared to the commercial multiplex kit. To overcome this undesired issue, we can improve the quality and robustness of our multiplex assay by improving the components of the reaction solution and the concentration of the primer-probe sets. Furthermore, our results showed that the multiplex assay is efficient for detecting all target genes (N, M1, NS1 and RNase P) in the same reaction tube (Co-infected samples). Finally the comparison between the two multiplex assays shows that our assay has advantages over the commercial kit, mainly including less PCR running time and higher detection sensitivity.

A limitation in this study was that we did not have access to the more validated clinical sample. Furthermore, some studies used a similar number of clinical samples. It should be noted that expanding the sample amount minimizes technical errors [23, 33]. Another limitation is that the study was conducted on frozen respiratory samples, which can lead to decrease in the sample quality.

Conclusion

In this study, a high-coverage multiplex RT-qPCR assay, with high analytical and diagnostic sensitivity and specificity was developed and validated. In a single reaction tube, it can facilitate the detection and differentiation of the SARS-CoV-2 (N gene), Influenza A (M1 gene), and Influenza B (NS1 gene), thereby increasing the testing throughput and further reducing both cost and time significantly. This multiplex assay will be useful during “Influenza seasons” when Influenza is expected co-circulate with SARS-CoV-2.

Data Availability

The authors confirm that all data generated or analyzed during this study are included in this published article [and/or] its supplementary information files.

References

  1. Fallah A, et al. Features of pathobiology and clinical translation of approved treatments for Coronavirus Disease 2019. Intervirology. 2022;65(3):119–33.

    Article  CAS  PubMed  Google Scholar 

  2. Hu B et al. Characteristics of SARS-CoV-2 and COVID-19 nature reviews microbiology. Nat Res, 2020.

  3. Mann R, et al. Clinical characteristics, diagnosis, and treatment of major coronavirus outbreaks. Front Med. 2020;7:581521.

    Article  Google Scholar 

  4. Bouvier NM, Palese P. The biology of Influenza viruses. Vaccine. 2008;26:D49–D53.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  5. Kumar B, et al. The emerging Influenza virus threat: status and new prospects for its therapy and control. Arch Virol. 2018;163(4):831–44.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  6. Frediansyah A, et al. Antivirals for COVID-19: a critical review. Clin Epidemiol Global Health. 2021;9:90–8.

    Article  CAS  Google Scholar 

  7. Swets MC, et al. SARS-CoV-2 co-infection with influenza viruses, respiratory syncytial virus, or adenoviruses. The Lancet. 2022;399(10334):1463–4.

    Article  CAS  Google Scholar 

  8. Hashemi SA, et al. High prevalence of SARS-CoV‐2 and Influenza A virus (H1N1) coinfection in dead patients in northeastern Iran. J Med Virol. 2021;93(2):1008–12.

    Article  CAS  PubMed  Google Scholar 

  9. Ma S, et al. Clinical characteristics of critically ill patients co-infected with SARS-CoV-2 and the Influenza virus in Wuhan, China. Int J Infect Dis. 2020;96:683–7.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  10. Havasi A, Influenza A, Influenza B et al. and SARS-CoV-2 similarities and differences–a focus on diagnosis. Front Microbiol, 2022: p. 2148.

  11. Abbasi H, et al. Real-time PCR ct value in SARS-CoV-2 detection: RdRp or N gene? Infection. 2022;50(2):537–40.

    Article  CAS  PubMed  Google Scholar 

  12. Chung H-Y, et al. Novel dual multiplex real-time RT-PCR assays for the rapid detection of SARS-CoV-2, Influenza A/B, and respiratory syncytial virus using the BD MAX open system. Volume 10. Emerging microbes & infections; 2021. pp. 161–6. 1.

  13. Pabbaraju K, et al. Development and validation of a multiplex reverse transcriptase-PCR assay for simultaneous testing of Influenza A, Influenza B and SARS-CoV-2. J Virol Methods. 2021;293:114151.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  14. https://www.idtdna.com/pages/education/decoded/article/calculations-converting-from-nanograms-to-copy-number.

  15. Ahmadi K, et al. Comparison of prevalence of Influenza and COVID-19: a Report from Hormozgan Province-Iran, 2019–2021. Disease and Diagnosis. 2022;11(4):131–6.

    Article  Google Scholar 

  16. Grohmann L et al. Guidance document on multiplex real-time PCR methods. 2021.

  17. Hirotsu Y, et al. Analysis of Covid-19 and non-covid-19 viruses, including influenza viruses, to determine the influence of intensive preventive measures in Japan. J Clin Virol. 2020;129:104543.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  18. Chung H-Y, et al. Multicenter study evaluating one multiplex RT-PCR assay to detect SARS-CoV-2, Influenza A/B, and respiratory syncytia virus using the LabTurbo AIO open platform: epidemiological features, automated sample-to-result, and high-throughput testing. Aging. 2021;13(23):24931.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  19. Rangaiah A, et al. Detection of SARS-CoV-2 in clinical samples: target-specific analysis of qualitative reverse transcription–polymerase chain reaction (RT-PCR) diagnostic kits. IJID Reg. 2021;1:163–9.

    Article  PubMed  PubMed Central  Google Scholar 

  20. Mentes A et al. Identification of mutations in SARS-CoV-2 PCR primer regions. 2022.

  21. Gdoura M, et al. SARS-CoV2 RT-PCR assays: in vitro comparison of 4 WHO approved protocols on clinical specimens and its implications for real laboratory practice through variant emergence. Virol J. 2022;19(1):1–9.

    Article  Google Scholar 

  22. Lesbon JCC, et al. Nucleocapsid (N) gene mutations of SARS-CoV-2 can affect real-time RT-PCR diagnostic and impact false-negative results. Viruses. 2021;13(12):2474.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  23. Tombuloglu H, et al. Development of multiplex real-time RT-PCR assay for the detection of SARS-CoV-2. PLoS ONE. 2021;16(4):e0250942.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  24. Sohni Y. Variation in LOD across SARS-CoV-2 assay systems: need for standardization. Oxford University Press US; 2021.

  25. Alcoba-Florez J, et al. Sensitivity of different RT-qPCR solutions for SARS-CoV-2 detection. Int J Infect Dis. 2020;99:190–2.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  26. Mboumba Bouassa R-S, et al. Analytical performances of the AMPLIQUICK® Respiratory Triplex assay for simultaneous detection and differentiation of SARS-CoV-2, Influenza A/B and respiratory syncytial viruses in respiratory specimens. PLoS ONE. 2022;17(1):e0262258.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  27. Kim M-J, et al. Evaluation of the AccuPower® RV1 Real-Time RT-PCR kit and the AccuPower® RV1 Multiplex Kit for SARS-CoV-2 and Influenza Virus Detection. Lab Med Online. 2021;11(4):290–6.

    Article  Google Scholar 

  28. Ni M, et al. Simultaneous detection and differentiation of SARS-CoV-2, Influenza A virus and influenza B virus by one-step quadruplex real-time RT-PCR in patients with clinical manifestations. Int J Infect Dis. 2021;103:517–24.

    Article  CAS  PubMed  Google Scholar 

  29. Cecilia D, et al. Development of a multiplex real-time RT-PCR assay for simultaneous detection of dengue and chikungunya viruses. Arch Virol. 2015;160(1):323–7.

    Article  CAS  PubMed  Google Scholar 

  30. Yun J et al. Evaluation of three Multiplex Real-time reverse transcription PCR assays for simultaneous detection of SARS-CoV-2, Influenza A/B, and respiratory syncytial virus in nasopharyngeal swabs. J Korean Med Sci, 2021. 36(48).

  31. Shu B, et al. Design and performance of the CDC real-time reverse transcriptase PCR swine flu panel for detection of 2009 A (H1N1) pandemic Influenza virus. J Clin Microbiol. 2011;49(7):2614–9.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  32. Shu B, et al. Multiplex real-time reverse transcription PCR for Influenza A virus, Influenza B virus, and severe acute respiratory syndrome coronavirus 2. Emerg Infect Dis. 2021;27(7):1821.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  33. Tombuloglu H, et al. Multiplex real-time RT-PCR method for the diagnosis of SARS-CoV-2 by targeting viral N, RdRP and human RP genes. Sci Rep. 2022;12(1):2853.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  34. World Health Organization. Available online: https://www.who.int/docs/default-source/coronaviruse/whoinhouseassays.pdf (accessed on 13 May 2020).

Download references

Acknowledgements

We are warmly grateful to all those who were involved in this study. We are thankful to Dr Mohammad Hassan Pouriayevali “COVID-19 National Reference Laboratory, Pasteur Institute of Iran” for helpful discussions. Also, we would like to thank Dr. Alijan Tabaraei and Naeme Javid for their technical assistance.

Funding

This study was supported by a grant from the Golestan University of Medical Science (Grant number: 111739).

Author information

Authors and Affiliations

Authors

Contributions

All authors give final approval of the manuscript to be submitted. Concept and design: Ayyoob KhosraviSupervision: Ayyoob Khosravi Drafting of the manuscript: Hamidreza Abbasi Administration, methodology, validation and investigation: Hamidreza Abbasi Submission of the manuscript: Hamidreza Abbasi and Hadi Razavi NikooTechnical or material support: Ayyoob Khosravi, Fatemeh Fotouhi and Hadi Razavi NikooFinal editing the manuscript: Ayyoob Khosravi, Fatemeh Fotouhi and Hadi Razavi NikooStatistical analysis: Hamidreza Abbasi.

Corresponding authors

Correspondence to Hadi Razavi Nikoo or Ayyoob Khosravi.

Ethics declarations

Ethics approval and consent to participate

This project is based on adhering to the declaration of Helsinki of 2013 and was approved by the Ethics Committee of Golestan University of Medical Science (Ethics code: IR.GOUMS.REC.1399.432). The covid-19 patients were not directly enrolled in this study. The covid-19 positive samples were prepared from the covid-19 bank of the Golestan University of Medical Science that was diagnosed and stored previously at -80°c.

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Electronic supplementary material

Below is the link to the electronic supplementary material.

Supplementary Material 1

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Abbasi, H., Nikoo, H.R., Fotouhi, F. et al. Development of a robust TaqMan probe-based one-step multiplex RT-qPCR for simultaneous detection of SARS-CoV-2 and Influenza A/B viruses. BMC Microbiol 23, 335 (2023). https://0-doi-org.brum.beds.ac.uk/10.1186/s12866-023-03048-9

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://0-doi-org.brum.beds.ac.uk/10.1186/s12866-023-03048-9

Keywords