Skip to main content

Table 4 Performance of Allplexâ„¢H pylori and ClariR Assay and DPO-PCR for detection of A2142G mutation compared with gene sequencing

From: Quantitative multiplex real-time polymerase chain reaction assay for the detection of Helicobacter pylori and clarithromycin resistance

Tests performances

Allplexâ„¢

DPO-PCR

Total positive results of the reference method (n)

2

2

Total positive results of the used test (n)

2

4

Total negative results of the used test (n)

140

138

False positive (n)

0

3

False negative (n)

0

1

Sensitivity %, (95% CI)

100 (15.8–100)

50.0 (1.3–98.7)

Specificity %, (95% CI)

100 (97.4–100)

97.9 (93.9–99.6)

Positive predictive value %, (95%CI)

100

25.0 (5.3–66.4)

Negative predictive value %, (95% CI)

100

99.3 (97.2–99.8)

Test Accuracy %, (95% CI)

100 (97.4–100)

97.2 (92.9–99.2)

Kappa (κ) (95% CI)†

1.00 (1.00–1.00)

0.62 (0.00-0.98)

  1. † Cohen’s kappa coefficient, Kappa values representing levels of agreement are categorized as follows: >0.90, almost perfect; 0.80 to 0.90, strong, 0.60 to 0.79, moderate; 0.40 to 0.59, week; 0.21 to 0.39, minimal; (0 to 0.20), none
  2. Abbreviations: Allplexâ„¢, Allplexâ„¢H pylori and ClariR Assay; DPO-PCR, dual priming oligonucleotide-based multiplex polymerase chain reaction; CI, confidence interval