Skip to main content

Table 2 Performance of the Allplexâ„¢H pylori and ClariR Assay and DPO-PCR in H. pylori detection compared with gene sequencing

From: Quantitative multiplex real-time polymerase chain reaction assay for the detection of Helicobacter pylori and clarithromycin resistance

 

No. (%) of results by reference standard

 

Analytical performance (%)‡

 

TP

TN

FP

FN

 

Sensitivity

Specificity

PPV

NPV

Accuracy

Kappa (κ)†

Allplexâ„¢

124 (98.3)

17 (12.0)

0 (0)

1 (0.7)

 

99.2 (95.6–100.0)

100 (80.5–100)

100

94.4 (70.7–99.2)

99.3 (96.1–100)

0.97 (0.91-1)

DPO-PCR

120 (84.5)

17 (12.0)

0 (0)

5 (3.5)

 

96.0 (90.9–98.7)

100 (80.5–100)

100

77.3 (59.0-88.9)

96.5 (92.0-98.9)

0.85 (0.72–0.98)

  1. † Cohen’s kappa coefficient, Kappa values representing levels of agreement are categorized as follows: >0.90, almost perfect; 0.80 to 0.90, strong, 0.60 to 0.79, moderate; 0.40 to 0.59, week; 0.21 to 0.39, minimal; (0 to 0.20), none
  2. ‡ Values in parentheses are 95% confidence intervals